K Logo Kamiya Biomedical Company
K-ASSAY
Since 1983
K-Assay
 
Insulin Control
Cat. No.: K73C-4M
Type: Control


KIT COMPOSITION
Cat. No. K73C-4M
Level 1 Human Insulin 2 x 0.5 mL, lyophilized
Level 2 Human Insulin 2 x 0.5 mL, lyophilized
   (Image coming soon.)
Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   Insulin Control is intended for use as consistent test sample of known concentration for monitoring the performance of the K-ASSAY ®   Insulin immunoturbidimetric assay.
FOR IN VITRO DIAGNOSTIC USE.

CONTROL PREPARATION
 1.  Allow controls to come to room temperature.  Remove cap carefully.

 2.  Add 0.5 mL of deionized water.  Let rest for 15 minutes.

 3.  Invert gently until dissolved.

STORAGE AND HANDLING
Store controls at 2-8°C.  After reconstitution, the controls are stable for 2 weeks if tightly capped and stored at 2-8°C. Unopened controls can be used for up to 18 months from the date of manufacture, as indicated by the expiration date on the package and bottle labels.

EXPECTED VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   Insulin Control are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert and/or vial labels for each lot for the expected values.

 REAGENT   LEVEL 1   LEVEL 2 
 MEAN 
 (µIU/mL) 
 RANGE 
 (µIU/mL) 
 MEAN 
 (µIU/mL) 
 RANGE 
 (µIU/mL) 
 K-ASSAY ®  Insulin   25    20 - 30   50   40 - 60