K Logo Kamiya Biomedical Company
K-ASSAY
Since 1983
K-Assay
 
CRP(3) Calibrator Set F
Cat. No.: KAI-086C
Type: Calibrator


KIT COMPOSITION
Calibrator 1-5 Human CRP 5 x 2 mL
    
KAI-086C
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   CRP(3) Calibrator Set F is intended to be used for the calibration of the K-ASSAY ®   CRP(3) immunoturbidimetric assay (Cat. No. KAI-082) for quantitating C-Reactive Protein (CRP) in serum and plasma samples (sodium EDTA or lithium heparin).
FOR IN VITRO DIAGNOSTIC USE.

CALIBRATOR PREPARATION
The calibrators are ready to use.  They do not require reconstitution.

STORAGE AND HANDLING
All calibrators should be stored refrigerated (2-8°C).  Return all calibrators to 2-8°C promptly after use.  Unopened calibrators can be used for 2 years from the date of manufacturing as indicated on the expiration date on the package and bottle labels.  Opened bottles of calibrators can be used for 1 month if stored at 2-8°C.

CALIBRATOR VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   CRP(3) Calibrator Set F are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert for each lot for the exact calibrator values.

  Cal #     mg/dL     mg/L     µg/dL  
  1     1     10     1,000  
  2     4     40     4,000  
  3     16     160     16,000  
  4     32     320     32,000  
  5     48     480     48,000  

Note: Isotonic saline (not included) should be used for Calibrator 0. (0 mg/dL)

This calibrator is traceable to IFCC CRM470.