KIT COMPOSITION
Lyophilized Human Serum 1 x 1 mL
Diluent (saline solution) 2 x 2 mL
    
KAI-018C
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   Lp(a) Calibrator is intended to be used for the multi-point calibration of the K-ASSAY ®   Lp(a) immunoturbidimetric assay (Cat. No. KAI-017 or KAI-044).
FOR IN VITRO DIAGNOSTIC USE.

CALIBRATOR PREPARATION
 1.  Add exactly 1 mL of purified water to the vial containing the lyophilized serum to dissolve it.

 2.  Prepare 5 calibrators (B-F) using the dissolved serum and diluent.  See product package insert for details.

STORAGE AND HANDLING
All calibrators should be stored refrigerated (2-8°C).  Return all calibrators to 2-8°C promptly after use.  Unopened calibrators can be used for one year from the date of manufacturing as indicated by the expiration date on the package and bottle labels.  Opened, reconstituted, and diluted calibrators can be used for 2 weeks if stored at 2-8°C.

CALIBRATOR VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   Lp(a) Calibrator are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert for each lot for the exact calibrator values.

Concentration of lyophilized serum, reconstitued with 1 mL of purified water (undiluted) :
Lp(a)
   156 mg/dL   

After reconstitution and dilution :
Calibrators A-F, values given in mg/dL
A B C D E F
  0.0     15.6     31.2     62.4     93.6     156.0