K Logo Kamiya Biomedical Company
K-ASSAY
Since 1983
K-Assay
 
Prealbumin Calibrator
Cat. No.: KAI-054C
Type: Calibrator


KIT COMPOSITION
Calibrator A 150 mM Sodium Chloride 1 x 1 mL
Calibrator B-F Human Prealbumin 5 x 1 mL
    
KAI-054C
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   Prealbumin Calibrator is intended to be used for the calibration of the K-ASSAY ®   Prealbumin immunoturbidimetric assay (Cat. No. KAI-053) for quantifying prealbumin in serum specimens.
FOR IN VITRO DIAGNOSTIC USE.

CALIBRATOR PREPARATION
The calibrators are ready to use and do not require reconstitution.

STORAGE AND HANDLING
All calibrators should be stored refrigerated (2-8°C).  Return all calibrators to 2-8°C promptly after use.  Unopened calibrators can be used for 24 months from the date of manufacturing as indicated on the expiration date on the package and bottle labels.  Opened bottles of calibrators can be used for 1 month if stored at 2-8°C.

CALIBRATOR VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   Prealbumin Calibrator are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert for each lot for the exact calibrator values.

  A     0 mg/dL  
  B     5 mg/dL  
  C     15 mg/dL  
  D     30 mg/dL  
  E     45 mg/dL  
  F     60 mg/dL  

The Prealbumin Calibrator values have been standardized against the IRMM ERM-DA470 reference material.