K Logo Kamiya Biomedical Company
K-ASSAY
Since 1983
K-Assay
 
Immunoassay Reagents
For Use on Automated Chemistry Analyzers
 
Krebs von den Lungen-6 (KL-6)
Cat. No.: KAI-250
Type: Reagent
Method: Immunoturbidimetric
Format: Liquid
Linearity: 70 - 10,000 U/mL
Sample: Serum, Plasma


KIT COMPOSITION
R1:   1 X 15 mL buffer
R2:   1 X 5 mL antibody

Tests / Kit:   100 (Hitachi 917)
(Can Vary Based On Analyzer Model)
    
KAI-250
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheets (SDS):
Reagent-1 (R-1)
Reagent-2 (R-2)

Click here for information about recommended calibrators and controls.


INTENDED USE
For the quantitative determination of Krebs von den Lungen-6 in human serum and plasma by immunoturbidimteric assay.
FOR RESEARCH USE ONLY IN THE U.S.
NOT FOR USE IN DIAGNOSTIC PROCEDURES IN THE U.S.

REAGENT PREPARATION
The reagents are ready to use.  They do not need to be reconstituted.

STORAGE AND HANDLING
All reagents should be stored refrigerated (2-8°C).  Return all reagents to 2-8°C promptly after use.  Unopened reagents can be used for 1 year from the date of manufacture, as indicated by the expiration date on the package and bottle labels.  Opened reagents can be used until the expiration date if they are kept tightly capped and at 2-8°C when not in use.

PERFORMANCE
Precision
With-run (Intra-assay) CV is less than 10% (n=5).

Accuracy / Correlation
A comparison of the K-ASSAY ®   KL-6 assay and another company's KL-6 latex-enchanced immunoturbidimetric assay was performed with the following results:

y = 1.065x - 2.3
r = 0.990
n = 122
x = another company's latex-enhanced turbidimetric KL-6 assay
y = K-ASSAY ®   KL-6 assay

An additional comparison with another company's KL-6 ECLIA (Electro Chemiluminesence Immuno-Assay) was performed with the following results:

y = 0.973x - 172.6
r = 0.957
n = 73
x = another company's ECLIA KL-6 assay
y = K-ASSAY ®   KL-6 assay

Assay Range
70 - 10,000 U/mL

INTERFERENCE
Bilirubin F :     No interference up to 18.9 mg/dL
Bilirubin C :     No interference up to 21.3 mg/dL
Chyle (Formazine Tubidity) :     No interference up to 1,410
Hemoglobin :     No interference up to 500 mg/dL
RF :     No interference up to 450 IU/mL
Sodium Citrate :     No interference up to 150 mg/mL
Sodium Heparin :     No interference up to 1,500 U/mL
EDTA-2Na :     No interference up to 15 mg/mL